Front Electrode Connection
- Use an air duster to remove and debris from the front electrode pads.
 - Prepare diluted conductive epoxy using the process described earlier.
 - Place the module assembly on the angled support plate, under a microscope.
 - Prepare a dispensing tool - attach a sewing needle to a long rod using tape.
 - Collect a drop of diluted conductive epoxy on the needle tip.
 - Place the drop on the PCB pad, and move the needle tip in small circles so that the drop touches the front electrode. The conductive epoxy should form a smooth fillet.
 - Repeat this process for all PZT elements in the transducer module.
 - Allow the conductive epoxy to cure for 24 hours.
 - Inspect the joints. Make sure that the conductive epoxy joints are flush with the edge of the PCB. If any of them overhang, use a new sharp scalpel blade to remove the excess material.
 